VOC Pro Transmitter • 53
NOTE:
Do not touch the new lamp window (the flat end) with your fingers as this may
contaminate the window with finger oil.
1. Press and hold the SUB button for approximately 6 seconds, until “OFF” shows on the
display screen.
2. Unscrew and remove the sensor housing cap from the sensor housing base. Set aside.
3. Gently unplug the sensor element from the sensor housing board.
4. Place the PID sensor face down on a flat clean working surface.
NOTE:
At this point it is recommended that the finger cots be used on the fingers handling the
lamp. Finger cots are included with the lamp cleaning kit.