5. COMMUNICATION DATA LIST
IMR02F04-E1
47
No. Name
RKC
Iden-
tifier
Modbus
register
address
Attri-
bute
Data range
Factory set
value
HEX
DEC
25 Input type
XI
00FA
250
R/W 0 to 4
Refer to
Input type and factory set
value table
.
Based on model
code.
26 Gain
setting
GA 00FB
251
R/W 0.500 to 4.000 mV/V or
0.5000 to 1.9999 mV/V
Varies with the setting of the Gain
setting decimal point position.
The setting value varies depending
on using pressure sensor.
When using our CZ-100P or
CZ-200P:
Refer to
Gain setting and
linearizing type (P. 53)
.
When using resistance for sensitivity
adjustment built-in pressure sensor:
Set the appropriate gain setting
value.
Refer to
Input type and
factory set
value table
27 Display
unit
PU 00FC
252
R/W 0: Kgf/cm
2
1: MPa
2: bar
3: psi
1
28 Input decimal point
position
XU 00FD
253
R/W 0: No decimal place
1: One decimal place
2: Two decimal places
3: Three decimal places
0
29 Pressure display high
XV
00FE
254
R/W Pressure display low to 19999
Varies with the setting of the Gain
setting decimal point position
50
30 Pressure display low
XW
00FF
255
R/W 0 to Pressure display high
Varies with the setting of the Gain
setting decimal point position
0
Input type and factory set value table
Set value
Input type
Factory set value
of gain setting (mV/V)
0
Our CZ-100P/CZ-200P (Standard)
1.500
1
Our CZ-100P/CZ-200P (Explosionproof)
1.500
2
Our CZ-100P/CZ-200P (Standard)
[Loose nut: 0.0 to 0.5 MPa, Fixed nut: 0 to 5 MPa]
0.650
3
Our CZ-100P/CZ-200P (Explosionproof)
[Loose nut: 0.0 to 0.5 MPa, Fixed nut: 0 to 5 MPa]
0.650
4
3.33 mV/V output type
(Pressure sensor made by other companies)
3.330