2. HOST COMMUNICATION
IMR02F04-E1
15
10. Set the Interval time.
Press the
key. The display goes to the
interval time.
Interval time
8n
1
InT
SET
Store
Press the
key to set the interval time.
Setting range: 0 to 250 ms
As an example “50” is set.
InT
001
0
Interval time
MONI
00
1
0
High-lighted digit
00
5
0
50 ms
(Four times)
MONI
High-
lighted
digit
Press the
key. The display goes to the set
lock level.
Set lock level
00
5
0
LCK
SET
Store
11. Go to the Measured value (PV).
Press and hold the
key for 2 seconds or
more until measured value (PV) is displayed.
Measured value (PV)
LCK
28
Set lock level
MONI
(2 seconds or more)
12. Enable communication parameter
When all communication parameter settings have
been completed, turn the power off and then on
to make the new set values take effect.
If you changed the communication
parameters, be sure to turn the power
OFF and then ON. If this is not done,
the higher level device will not be able to
recognize the changed values and
communication may not be possible.