Calibration
© Rice Lake Weighing Systems
●
All Rights Reserved
41
5.1.2 Linear Calibration
Linear calibration points provide increased scale accuracy by calibrating the indicator at up to four additional points between the
zero and span calibrations.
1. Complete
. Press
.
wlin
displays.
2. Press
.
point1
displays.
3. Press
.
wlin v1
displays.
4. Press
. The current test weight value for point 1 displays.
5. Press
and use the numeric keypad to enter a new value, if necessary.
6. Press
to accept value.
wlin c1
displays.
7. Place the specified amount of test weight on the scale.
8. Press
to perform a linear point calibration.
ok
displays.
9. Press
.
wlin f1
for more information on WLIN F#.
10. Press
.
Point1
displays.
11. Press
.
Point 2
displays.
12. Repeat previous steps for points 2-4, if necessary.
The linear calibration for a point is saved once point is calibrated.
13. Press
to return to weigh mode.
5.2 Alternative Zero Calibrations
During a calibration, the zero value (
wzero
) can be replaced with a temporary zero (
tmpzero
) or last zero (
lstzero
). A
rezero (
rezero
) can be done after calibration. See below for information on these alternative zeros.
5.2.1 Last Zero
This takes the last push-button zero in the system (from weigh mode) and uses it as the new zero reference point, after which a
new span calibration must be performed. This calibration cannot be performed when calibrating a scale for the first time.
A last zero calibration is typically used on truck scales to allow a scale verification to become a calibration without having to
remove the test weights.
5.2.2 Temporary Zero
A temporary zero calibration temporarily zeros the displayed weight of a non-empty scale. After span calibration, the difference
between the temporary zero and the previously calibrated zero value is used as an offset.
A temporary zero calibration is typically used on hopper scales to calibrate the span without losing the original zero calibration.
5.2.3 Rezero
A rezero calibration is needed to remove a calibration offset when hooks or chains are required to suspend the test weights.
Once a span calibration is complete, remove the hooks or chains and the test weights from the scale. With all the weight
removed, a rezero calibration is used to adjust the zero and span calibration values.
Note
Содержание Revolution 680 Synergy Plus
Страница 1: ...680 Synergy Plus Digital Weight Indicator Technical Manual August 14 2019 PN 192627 Rev A...
Страница 79: ......