680 Synergy Plus – Digital Weight Indicator
26
Visit our website
4.4.2 Setup – Format Menu
4.4.2.1 Primary and Secondary Menus
Figure 4-6. Format – Primary and Secondary Menus
4.4.3 Setup – Calibration Menu
Figure 4-7. Setup – Calibration Menu
Parameter
Description
format
Format – Sets the decimal point and the display divisions for the Primary (
primar
) and Secondary (
secndr
)
display weight format; For example, select
8888.885
if a count by of 0.005 is needed or select
8888820
if a count by
of 20 is needed (the
8
s serve as placeholders and show a breakdown of how digits will display);
Settings:
8888881
(Primary default), 8888882, 8888885, 8888810, 8888820, 8888850, 8888100, 8888200,
8888500, 8.888881, 8.888882, 8.8888885, 88.88881, 88.88882, 88.88885, 888.8881, 888.8882, 888.8885,
8888.881, 8888.882, 8888.885, 88888.81, 88888.82, 88888.85, 888888.1, 888888.2, 888888.5 (Secondary default)
Units
Units – Sets the units type;
Settings:
LB
(Primary default), KG (Secondary default), OZ, TN, T, G, NONE
Enabled
Enabled – Enables the front panel UNITS button to toggle between the primary and secondary formats (only
displays under Secondary);
Settings:
ON
(default), OFF
Table 4-5. Format – Primary and Secondary Menu Parameters
Parameter
Description
wzero
Zero Calibration – Executes the zero calibration process; See
zerocnt
Zero Calibration Count – Displays the raw count value at the zero weight; A zero calibration (WZERO)
generates this raw count value; Manually changing this count value changes the zero weight and negates the
zero calibration
wval
Test Weight Value – Sets the weight value for the span calibration; See
;
Enter value: 0.000001–9999999.999999,
10000.0
(default)
wspan
Span Calibration – Executes the span calibration process; See
spancnt
Span Calibration Count – Displays the raw count value at the span weight; A span calibration (WSPAN)
generates this raw count value; Manually changing this count value changes the span weight and negates the
span calibration
Table 4-6. Setup – Calibration Menu Parameter
Setup
Format
FORMAT
UNITS
PRIMAR
FORMAT
UNITS
ENABLED
SECNDR
Setup
Calibr
zerocnt
WZERO
WVAL
spancnt
WSPAN
Point 1-4
WLIN
REZERO
Wlin v1-4
wlin c1-4
wlin f1-4
lstzero
tmpzero