Installation
© Rice Lake Weighing Systems
●
All Rights Reserved
11
2.7 Sealing the Indicator (Optional)
Insert a lead wire seal through three fillister screws. This restricts access of the setup switch, electronics, electrical contacts and
Legal for Trade configuration parameters.
The audit jumper (J24) must be set to off to require pressing the setup switch for access to configuration parameters.
Figure 2-10. Sealing the Indicator – No Access
1. Reposition the two fillister head backplate screws to the lower right and lower right-center screws locations.
1. Torque the two backplate screws and setup screw as specified in
2. Navigate the sealing wire through the fillister head screws on the backplate and the fillister head screw at the bottom of
the enclosure, as shown in
3. Seal the wire to secure.
2.8 Parts Kit Components
The recommended cable strip length is 0.25'' (7 mm) for all 680 connectors.
Part No.
Description
Qty
15631
Cable Tie, 3'' Nylon
4
15650
Mount, Cable Tie 3/4''
2
193230
Screw, Mach M4 x 0.7 x 10 Phillips Pan Head SST
4
194219
Screw, Mach M4 x 0.7 x 10 Phillips Drilled Cheese Head SST
2
194446
Ferrite Core, Snap on Fair-rite
1
194488
Screw, Mach M4 x 0.7 x 6 Phillips Pan Head with External Tooth Washer SEMS
4
19538
Post Plug, Slotted Black Plastic Stem, 1/4 x 1, Seals Inside Cord Grip
3
195993
Connector, 6 Position Screw Terminal Pluggable 3.50 mm Black
3
195995
Connector, 4 Position Screw Terminal Pluggable 3.50 mm Black
1
195998
Connector, 5 Position Screw Terminal Pluggable 3.50 mm Black
1
30632
Bag, Plastic 5 x 8
1
42149
Bumper, Rubber Grommet 0.50 (OD) x 0.281 (ID)
4
53075
Clamp, Ground Cable Shield, Radius 0.078''
4
75062
Washer, Bonded Sealing #8 7/16 (0.4375) OD SST
6
94422
Label, Capacity 0.40 x 5.00
1
Table 2-10. Parts Kit (PN 194477) Components
Note
Note