MSI TranSend 7000 Series
30
V
isit our website
www.ricelake.com
3.2.6 Inputs / Outputs
Setpoints
Setpoints provide a trip point for load values.
Figure 3-24. Setpoint Configuration
1. Select
Inputs / Outputs
, then
Setpoints
.
2. Enter the following parameters:
Setpoint Number
– select the setpoint number from 1-3
Status
– select Disabled or Enabled
Source Sensor ID
– select the source sensor ID number from 1-5
Relay Output Mode
– select Coil or Latch
Comparison Logic
– select Undefined, Greater Than or Less Than
Comparison Value
– enter the comparison value
Value Type
– select the value type parameter
Figure 3-25. Value Type Selections
Hysteresis in D
– select the hysteresis in D number from 0-99
3. Press
to save.
Stream Print String
See
Section 3.3.7 on page 37
for the advanced setup of the stream print string settings.
Do not change these settings without consulting Rice Lake Weighing Systems or a local dealer.
3.2.7 Monitors
Select the view to be used for viewing the data from the product.
Terminal
Streams the data in a print format.
Meter
Displays a meter view for viewing and troubleshooting.