ScaleCore Connect
©
Rice Lake Weighing Systems
●
All Rights Reserved
25
3.2.4 Load Cells
The general load cell display allows the load cells’ parameters to be set to for each load cell connected to the connected device.
Figure 3-16. Load Cell General Settings
1. Select
Load Cells
, then
General
.
2. Set the following general parameters as needed for the current product:
Load Cell Number
– select a load cell number 1-5
Load Cell Enabled
– select Enabled (default) or Disabled
Filter
– set filtering to Off, Low, Medium or High
Load Cell Name
– enter a name to identify the load cell
Motion Enabled
– select Enabled or Disabled (default)
Motion Detect Period in 50 mSec Tick
– select a number from 1-255
Motion Band In D, 0=0.5D, 1=1D etc.
– select a number from 1-255
Pending Time in 50 mSec Tick
– select a number from 1-255
Viewing Capacity
– enter a capacity
Viewing Countby
– select 0.0001-5000
Under Load Threshold
– select a number from -100 to 90
3. Press
to save.
Zero Standard
Figure 3-17. Load Cell Zero Settings
1. Select
Load Cells
, then
Zero, Standard
.