MSI TranSend 7000 Series
28
V
isit our website
www.ricelake.com
3.2.5 Calibration
A visual interface for performing the available types of load cell calibration to ensure the
ScaleCore
product is accurate.
The available methods of calibration are:
•
Full Cal Load Cell
– enter scale unit, capacity and countby
•
Re-Cal Load Cell
– uses current scale unit, capacity and countby
•
Full C-Cal Load Cell
– enter scale unit, capacity and countby; allows a calibration using a Constant Calibration (C-Cal)
number without the requirement of test weights
•
Re-C-Cal Load Cell
– use current scale unit, capacity and countby; allows a calibration using a C-Cal number without the
requirement of test weights
•
Multi Load Cell
– use when calibrating multiple load cells, perform a Full Calibration on each load cell to be calibrated
Full Calibration
1. Select
Calibration
, then
Full Cal Load Cell
.
Figure 3-23. Select Load Cell ID
2. Select the
Load Cell ID
number from 1-4.
3. Press
.
4. Set the
Unit
and
Capacity
parameters.
5. Press
.
Press
to end current calibration and restore the previous calibration.
6. Select the desired
Countby
parameter, then press
.
7. Ensure there is no weight on the load cell, then press
.
8. Enter the test weight value, then press
.
9. Press
. The constant cal number is displayed, document the number for use later if needed.
10. Repeat steps above for each load cell to be calibrated.
Re-Calibration
1. Select
Calibration
, then
Re-Cal Load Cell
.
2. Select the Load Cell ID number from 1-4.
3. Press
.
4. Ensure there is no weight on the load cell, then press
.
5. Enter the test weight value, then press
.
6. Press
. The constant calibration number is displayed, document the number for use later if needed.
7. Repeat steps above for each load cell to be re-calibrated.
Note