Installation
© Rice Lake Weighing Systems
●
All Rights Reserved
11
2.2.2 MSI-7000 TranSend and MSI-7000HD
For MSI-7000 TranSend and MSI-7000HD transmitters, feed standard load cell cables through the cord grips and wire to the
terminal strip.
Figure 2-2. Load Cell to Terminal Block Connection
In most applications it is not necessary to connect the shield to the j-box shield terminal. Sometimes noise pickup is
improved by connecting the shield inside the j-box. However, it may make noise pickup worse. Test performance
before cutting off the shield at the cable end.
2.2.3 MSI-7001HD
For MSI-7001HD transmitters, feed standard load cell cables through the cord grips and wire to the terminal strip.
Load Cells 1 and 2 share a common excitation connection in two load cell configurations.
Figure 2-3. MSI-7001HD Load Cell to Terminal Block Connection
Load Cell 1
Load Cell 2
Load Cell 3
Load Cell 4
– E
xc
ita
tio
n
+ E
xc
ita
tio
n
– S
ig
na
l
+ S
ig
na
l
– E
xc
ita
tio
n
+ E
xc
ita
tio
n
– S
ig
na
l
+ S
ig
na
l
– E
xc
ita
tio
n
+ E
xc
ita
tio
n
– S
ig
na
l
+ S
ig
na
l
– E
xc
ita
tio
n
+ E
xc
ita
tio
n
– S
ig
na
l
+ S
ig
na
l
Note
Note
Load Cell 2
Load Cell 1
– Signal 2
+ Signal 2
– Excitation
+ Excitation
– Signal 1
+ Signal 1
Shield
Shared
Содержание MSI-8000
Страница 49: ......