IQ9500 Counting Scale Setup
27
7.0
IQ9500 Counting Scale Setup
The following paragraphs contain instructions for
initial IQ9500 setup and operation either in the battery
or the AC power mode. These instructions include
setting the specification data bits for both internal and
external scale configurations.
7.1
Setting Up the IQ9500
When setting up the IQ9500 counting scale or adding
a second channel, third channel or a force balance, it
is necessary to assign a different location for each
scale whether it is present or not.
Specifications 16 and 17 are used only to set up the
annunciators for each of the four scales. These
specifications do not turn on the additional scales.
Scales 2 and 3 are turned on and off in the weights and
measures specifications contained in Specification 25,
bit 1 (
Scale Type
) and Specification 32, bit 0 (
Scale
Connected to the Cable Port 2 Connector
). The
force balance is turned on and off in Specification 8,
bit 3 (
RS-232C Connection, Force Balance
) and the
connector type is selected in Specification 13, bits 2,
1, and 0 (
RS-232 Connector
).
Table 7-1 contains the default bit configurations for
Specifications 16 and 17.
NOTE:
Specifications 16 and 17 can only be changed by
an authorized service technician.
Do not set Specifications 16 and 17 to the configuration shown below.
A lockup condition will result if
these configurations are selected. Contact an authorized service technician if a lockout condition
occurs.
Should the IQ9500 become locked up in all
888888
s
as a result of invalid specification settings, an
authorized service technician can only perform the
following procedure:
1. Unplug scale from AC outlet.
2. Press and hold the
3
and
9
keys while
plugging in AC cord to outlet.
3. The display will show
S-on
. Release the
3
and
9
keys.
4. Press and hold
REZERO
key while pressing
141
, press
+
key to advance to SPEC 16.
Change SPEC 16 and SPEC 17 to assign a
different location for each scale whether it is
present or not.
Specification
Default Setting
Bits 2 and 3
Bits 0 and 1
16
0001
Scale 1
00:
Internal Scale 1
01:
Internal Scale 2
10:
External scale
11:
Force balance
Scale 2
00:
Internal Scale 1
01:
Internal Scale 2
10:
External scale
11:
Force balance
17
1011
Scale 3
00:
Internal Scale 1
01:
Internal Scale 2
10:
External scale
11:
Force balance
Scale 4
00:
Internal Scale 1
01:
Internal Scale 2
10:
External scale
11:
Force balance
Table 7-1. Specifications 16 and 17 Default Settings
Specification
Default Setting
Bits 2 and 3
Bits 0 and 1
16
0010
Scale 1
00:
Internal Scale 1
Scale 2
10:
External scale
17
1000
Scale 3
10:
External scale
Scale 4
00:
Force balance
!
Caution