Calibration
25
6.0
Calibration
The calibration procedure maintains the scale
accuracy within specifications and can serve as a
performance test procedure. Your IQ9500 scale should
be turned on for a minimum of 10 minutes and the
platform excercised three to four times before
attempting to calibrate.
NOTE:
Use the
SCALE
key to select the platform to
calibrate. The scale number is displayed on the lower
secondary display. For dual-platform scales the
calibration procedure described below must be repeated
for each scale.
6.1
IQ9500 Display Resolution
The following tables list the weight resolution
capabilities for the IQ9500 counting scale. Table 6-1
has the single channel platform configuration while
Table 6-2 contains the dual channel scales. Use these
tables to verify specification settings before
calibration.
Capacity, lb
Weighing
Resolution
Specification 20
Specification 21
Specification 23
Specification 25
Specification 32
0.5
0.0001
0101
0100
0100
0000
1010
1
0.0001
0101
0100
0000
0000
1010
2.5
0.0002
0001
0100
0000
0000
1010
5
0.0005
1001
0100
0000
0000
1010
10
0.001
0101
0011
0000
0000
1010
25
0.002
0001
0011
0000
0000
1010
50
0.005
1001
0011
0000
0000
1010
100
0.01
0101
0010
0000
0000
1010
Table 6-1. Single Channel Platform
Capacity, lb
Weighing
Resolution
Specification
20
Specification
21
Specification
22
Specification
23
Specification
25
Specification
32
0.5/10
0.0001/0.002
0100
0100
0011
0100
0010
1010
0.5/25
0.0001/0.005
0110
0100
0011
0100
0010
1010
0.5/50
0.0001/0.01
0101
0100
0010
0100
0010
1010
1.0/10
0.0001/0.001
0101
0100
0011
0000
0010
1010
1.0/25
0.0001/0.002
0100
0100
0011
0000
0010
1010
1.0/50
0.0001/0.005
0110
0100
0011
0000
0010
1010
2.5/10
0.0002/0.001
0001
0100
0011
0000
0010
1010
2.5/25
0.0002/0.002
0000
0100
0011
0000
0010
1010
2.5/50
0.0002/0.005
0010
0100
0011
0000
0010
1010
5.0/10
0.0005/0.001
1001
0100
0011
0000
0010
1010
5.0/25
0.0005/0.002
1000
0100
0011
0000
0010
1010
5.0/50
0.0005/0.005
1010
0100
0011
0000
0010
1010
Table 6-2. Dual Channel Scales