BiliChek Service Manual
12
1014988
BiliCal
Individual
Calibration Tip
The green light will illuminate when the battery pack has
completely discharged and recharged. After both battery packs
have been discharged, place the Deep Discharge switch in the
left ( ) position.
Caution:
Leaving the Deep Discharge switch in the right
( ) position at all times will drain the battery pack before
every
recharge. This will increase the duration of the recharg-
ing cycle and may reduce the overall life of the battery packs.
NOTE:
If the battery pack is moved or bumped during the
recharge phase of the deep discharge cycle, the deep discharge
cycle may begin again.
The Battery Status Indicator (BSI)
BSI Color Battery Charger Mode
Solutions
OFF
Green
Amber
Amber Flashing
Red Flashing
Battery pack not sensed in
well or charger base not
properly connected to power
supply.
Battery pack is fully charged.
Battery pack in the well is
charging but not ready for use.
Battery pack in the well is
deep discharging.
Bad battery pack in the
charger base.
Ensure battery pack is
properly seated in
well, check power
connections. Try
another battery if BSI
is still off. If BSI is
still off with second
battery, the Charger
base itself has
malfunctioned and
requires service.
Battery pack is ready
for use.
Continue charging
until BSI color is
green.
Continue discharging
and recharging until
BSI color is green.
NOTE:
BSI will
become steady amber
when deep discharge
phase of cycle is
ended and charging is
in progress.
Replace the battery
pack. If the problem
is still present with a
new battery pack, the
charger base itself has
malfunctioned and
requires service.
In order to maintain optimal battery performance and battery
life, the battery packs should periodically be fully drained of
power. The charger base has the ability to manually drain the
battery by using the Charge/Deep Discharge switch located
on the bottom of the charger base. It is recommended by the
manufacturer to discharge each battery pack every month. To
drain a battery pack of power, place the Deep Discharge switch
in the right ( ) position and place the battery pack in the
charger base. The discharge/recharge cycle will take approxi-
mately 4.6 to 9.2 hours depending on the amount of the charge
in the battery pack.