BiliChek Service Manual
30
1014988
Error: E10 - No Lamp Current
Measurement lamp burned out. Replace measurement lamp and perform several calibrations until HHU calibrates. If HHU
will not calibrate after numerous tries, HHU requires service.
Error: E99 - Fatal
Fatal software error. The HHU must be returned for service. Contact your distributer to request instructions to return to
Respironics for repair.
The following tables are for information only. If an E99 occurs, record the sub error and place a note with the HHU when
returning it for service.
Error: E99 - Fatal (Sub Error Tables)
Sub Errors
Description
0001
System configuration parameters are not valid
0002
Timer could not be allocated
0004
Audio could not be initialized
0008
RTC could not be initialized
0016
Serial port could not be initialized
0032
LCD could not be initialized
0064
Switches could not be initialized
0128
Microspectrometer configuration data is not valid
0256
Internal reference voltage is out of range
0512
System RAM check failed
1024
Checksum error in boot firmware
2048
Checksum error in main firmware
4096
Firmware version in boot block not found
8912
Watchdog timer could not be initialized
16384
Watchdog timer timed-out
HHU Error Code list with Explanations (continued)
Содержание bilichek
Страница 1: ...Service Manual www respironics com ...
Страница 23: ...BiliChek Service Manual 23 1014988 11 3 3 Charger Base PCB Schematic ...
Страница 27: ...BiliChek Service Manual 27 1014988 11 3 4 Charger Base PCB Component Layout ...
Страница 36: ...Reorder B800 25 Revision 1014988 DS 07 14 03 www respironics com ...