NOTE:
The following sub errors are displayed in the date field.
Firmware version 2.3/3.6 and below
Sub Errors
Description
0256
External reference voltage outside of limits
0512
5V analog supply below limit
1024
Programming voltage below limit
2048
Clinical Error
4096
Internal system error detected during clinical testing
8192
Microspectrometer SN = 000000
Firmware version 2.4/3.8 and above
Sub Errors
Description
0016
Peak pixel range > 25
0032
External reference voltage outside of limits
0064
5V analog supply below limit
0128
Programming voltage below limit
0256
Clinical error
0512
Internal system error detected during clinical testing
1024
Microspectrometer SN = 000000