BiliChek Service Manual
29
1014988
Error: E02 - Measurement
Sub Errors Description
000001
Light intensity below lower threshold specified in data configuration parameters. Try several calibrations
and measurements. If error persists, try on another patient. If error persists on several patients, HHU
requires service.
000002
Light intensity above upper threshold specified in data configuration parameters. Same as in 000001.
000008
Bili
Chek
bilirubin measurement below lower limit. Same as in 000001.
008192
Stray Light Slope Error. Try several calibrations and several Bili
Cal
tips. If error persists, HHU requires
service.
016384
Mean of dark scan data above threshold. Try several calibrations and measurements. If error persists,
HHU requires service.
Error: E03 - Disposable
Disposable tip removed after calibration and before measurements completed. Ensure tip is firmly attached an try another
calibration and measurement. NOTE: This error can also occur if HHU is operated in sunlight or other bright light. Cover the
IRDA port on the front of the HHU and perform another calibration and another measurement.
Error: E04 - Movement
Excessive movement detected during measurement. Take another measurement ensuring no movement of the Bili
Chek
hand-
held unit while the MSI is green.
Error: E05 - Bili
Chek
Bilirubin Measurement High
Bili
Chek
Bilirubin measurement is above the upper threshold. The Bili
Chek
Non-Invasive Bilirubin Analyzer has an upper
threshold of 20 mg/dl or 340
µ
mol/L set in the software. If readings are above this threshold, a blood sample must be
taken to measure the TSB level.
NOTE: Firware
Version
≥
2.4/3.8
Only