11.0 Troubleshooting Guide
Display Window
Indication
Solution
E00
No user set-up information identified.
Install a charged battery pack into the HHU. Press
F1
and
F2
at the same time, hold for one second and then release. This
places the HHU in the Set-up mode. Enter settings specific to your
institution. See section 10.0 of this manual for Set-up Mode
instructions.
E01
Calibration measurement is unacceptable.
Press and release
F1
and
F2
at the same time.
Ensure Bili
Cal
tip is in place and retry the calibration. If the error
continues: 1. Place a
fully charged
Battery Pack in the HHU, install a
new BiliCal Tip, and try again to calibrate several times.
2. Remove and rotate the Bili
Cal
tip and retry. 3. Replace the measurement
light bulb (see Section 14.1 for instructions) and retry. If the error is still
present, the HHU has malfunctioned and requires service. Call
Respironics or your distributor.
E02
Patient measurement is unacceptable.
Press and release
F1
and
F2
at the same time. Take another
measurement. If E02 occurs again, replace the Bili
Cal
tip, repeat
calibration and retry the measurement. If the error is still present, the
HHU has malfunctioned and requires service. Call
Respironics or your distributor.
E03
Bili
Cal
Individual Calibration Tip removed
Press and release
F1
and
F2
at the same time.
between calibration and measurement scan.
Remove the Bili
Cal
Individual Calibration Tip.
Replace with a new tip and recalibrate.
E04
Too much movement during measurement.
Press and release
F1
and
F2
at the same time.
Perform another measurement.
E05
High Measurement is out of clinically proven range.
Obtain a Total Serum Bilirubin level (TSB)
E10
Light bulb is burned out.
Replace measurement light bulb (see section 14.1 for instructions). If
error continues, the HHU has malfunctioned and requires service. Call
Respironics or your distributor.
E99
System error is detected.
Call for Service.
Low battery pack. Maximum of 10 measurements left.
Replace current battery pack with a fully
charged battery pack from the charger base and place the discharged
battery in well B for charging.
Discharged Battery pack.
Replace current battery pack with a fully charged one from the charger
No measurements can be taken.
base. Place the discharged battery pack into well B for charging.
Display
Window
Indication
Solution
11.1 Troubleshooting Guide
Содержание bilichek
Страница 1: ...Service Manual www respironics com ...
Страница 23: ...BiliChek Service Manual 23 1014988 11 3 3 Charger Base PCB Schematic ...
Страница 27: ...BiliChek Service Manual 27 1014988 11 3 4 Charger Base PCB Component Layout ...
Страница 36: ...Reorder B800 25 Revision 1014988 DS 07 14 03 www respironics com ...