XCell™ Lab Controller
User Guide
31 repligen.com XC-LAB-UG-V4
Table 12. Starting, pausing and stopping XCell ATF® Device(s) in Dual Mode
Actions
In-phase
and out-of-
phase
Start
Open either the ATF-A or ATF-B flow Pop-
up , select mode for dual control and then
press
Start.
Pause
1.
Software: Open either ATF -A or ATF-
B Flow Pop-up and press
Pause
2.
Software: Press the
All Pause
buttons
on ATF Main screen
3.
Hardware: Press either
Pause
button
on controller Face B (Figure 6)
Stop
Software: Press the
All Pause
buttons on
ATF Main screen
Start
1. Software: Open either ATF -A or ATF-B
Flow Pop-up and press
Pause
2. Hardware: Press either
Pause
button on
controller face B
Independent
Start
Open the relevant ATF flow Pop-up and
then press
Start.
Pause
4.
Software: Open either ATF -A or ATF-
B Flow Pop-up and press
Pause
5.
Software: Press the
All Pause
buttons
on ATF Main screenn to pause the
respective ATF unit (A or B) to pause.
For example if the all pause was
pressed for ATF B while running in
independent mode, then ATF B is
paused and ATF A continues to run.
Hardware: Press either
Pause
button on
controller Face B (
Stop
Press
Stop
on the
ATF Flow pop-up.
Upon
re-start, the controller starts without
utilizing previous settings. A
Stop
is
recommended to be used only at the end
of a run.
Note:
It is not possible to re-start in dual mode unless both ATFs are paused and re-started together.
Note:
After pressing STOP, the setpoint value remains the same, which is the only user editable
parameter. The most significant difference between STOP and PAUSE is that PAUSE keeps the
XCell ATF® Device in a hold position where it can resume when START is pressed a second time.
Pressing STOP stops the running session; when start is subsequently pressed it goes through
the start initiation process (Priming).