XCell™ Lab Controller
User Guide
18 repligen.com XC-LAB-UG-V4
Table 5. Controller ports
Item
Description
1
Air
Positive air pressure from SAPA
2
Vacuum
Vacuum line from facility line or vacuum pump manifold
3
Multi-pumps
NOT CURRENTLY USED
4
Permeate A pump
NOT CURRENTLY USED
5
Permeate B pump
NOT CURRENTLY USED
6
AUX Pumps
NOT CURRENTLY USED
7
Power switch
Power on/off
8
24V DC in
DC power from power supply
9
24V DC out
NOT CURRENTLY USED
10 ATF A / B System pause
Pause buttons for ATF A and ATF B
LED light to indicate status
11 Alarm
Visual alarm (audio optional)
12 To ATF A / B
A2C connections for air pressure and vacuum to ATF
filtration devices
13 Bioreactor A / B Weight
NOT CURRENTLY USED
14 ATF A / B Flow Sensor
A2B retentate line flow sensor connections from filtration
devices
15 Ethernet
1: Communication between controller and tablet (M12 via
ethernet to RJ45 USB adapter on tablet)
2:
NOT CURRENTLY USED
16 ATF A / B Permeate pressure
Permeate line P3 pressure sensor inlet connections
17 VCD – Bioreactor A / B
NOT CURRENTLY USED
Two manual pause buttons (item 10) allow each device to be independently paused. The alarm (11)
alerts the user when attention is required. The A2C lines (12) provide positive and negative air
pressure to the XCell ATF® Device. Pressure (16) and flow (14) sensor ports carry data from the
device-mounted sensor to the controller. An ethernet port (15) transfers data to a tablet for
viewing, processing, and analysis. Like Face A, some ports of Face B are not used in the current
model.
The software interface is accessed through a Surface Pro tablet, provided in a protective case. The
tablet may be mounted directly on the controller or to a shelf, using a bracket and moveable support
arm. Each controller requires a dedicated tablet that uses a wired ethernet connection to the
controller port.
6.5.1
A2B and A2C tubing sets
Each XCell ATF® Device requires a specific A2B and A2C tubing set size. The controller ships with a
collection of A2C tubing sets suitable for all devices (ATF1, ATF2 and ATF4). The A2C tubing may
appear to be similar – but they are not – and each is designed to function with a specific XCell ATF®
device size. Labeling on the tube sets indicates use with an ATF1, ATF2 or ATF4 device. The A2C
tubing set connects the XCell ATF® Device to the controller.
A flow sensor on the A2B line, engineered specifically for XCell ATF® applications, requires a specific
tubing OD and tubing type. Tube sets with a permeate pressure sensor also are available. Both the
flow sensor and the pressure sensor connect directly to the controller. The device specific setup-
guides describe tube set specifications in detail; a brief summary is provided in the table below.