34
Interconnection schemes
6.4 Preparation of integration kits
When the
SPA
2
integration kit is received it will contain the necessary components required connect the
SPA
2
and either the
UCC
2
or
UCC
lite
controller system to a CMM installation.
NOTE:
All connectors used in the integration of the
SPA
2
to a CMM installation are widely commercially
available, the connector kits are designed to ease the system integration by permitting the connectors to
be purchased at the same time as the
SPA
2
.
6.4.1 Motor
connectors
Each connector kit will contain three motor connectors which are appropriate for DC motor output from the
SPA
2
.
NOTE:
This connector is of the solder bucket variety, it is recommended that it is fitted to the CMM wiring
by a competent technician or engineer, and that sleeving is used for all connections to reduce the
possibility of short circuits.
Each of the three connector assemblies consist of the following:
•
1 off metal backshell
•
1 off 7W2 power and signal D-type connector
•
2 off power pins
Содержание apply innovation SPA2