18
Rear panel description
3.3.6 Earth
stud
The
SPA
2
has an earth stud to permit equipotential bonding of the CMM control system to that of the rest
of the CMM.
NOTE:
When the
SPA
2
is used with the
UCC
2
or
UCC
lite
controller the CMM frame should be bonded to
the
SPA
2
earth stud and not the
UCC
2
or
UCC
lite
earth stud.
3.3.7
External emergency stop connection
This 9-pin D-type connection is designed to permit connection of external emergency stop devices to the
Renishaw controller emergency stop system.
Emergency stop devices must meet the requirements of IEC 60947-5-1 (low-voltage switchgear and
controlgear – part 5-1: control circuit devices and switching elements – electromechanical control circuit
devices) or UL1054 (standard for special-use switches).
The machine manufacturer or product installer must perform a risk assessment to determine the
requirements for emergency stopping and emergency switching off. The
SPA
2
emergency stop safety
system is designed to achieve category 2 to standard EN954-1:1996 (ISO 13849-1). The risk assessment
should therefore have determined that a category 2 safety function is satisfactory.
There are three connection pins available on this connector. Please refer to the table below for their
functions:
Pin number
Function
6
ESTOP A
7
ESTOP B
9
RESET B
Any additional emergency stop component that is fitted to this connector must have the following electrical
characteristics:
Emergency stop system voltage : 24 V
Emergency stop system current : 1 A
NOTE:
If the emergency stop reset function is not required then it will be necessary to connect pins 5 and
9 so that when an emergency stop is removed from the system, it will automatically reset.
If no additional emergency stop devices are to be added to a
UCC
2
or
UCC
lite
controller system, connect
pins 6 and 7 to permit the MCU1 emergency stop switch to function.
It is strongly recommended that all connections to this connector be fitted by a competent technician or
engineer and that all wires should be sleeved. See also 6.4.2.
It should not be necessary to switch off the
SPA
2
in an emergency but if a requirement is indicated by the
user’s risk assessment, emergency switching off for the complete machine must be implemented
externally to the
SPA
2
.
Содержание apply innovation SPA2
Страница 1: ...Installation guide H 1000 5234 03 A SPA2 servo power amplifier...
Страница 3: ...SPA2 Installation guide...
Страница 15: ...System description 13 1 2 7 SPA2 and UCC2 system layout 1 2 8 SPA2 and UCClite...
Страница 34: ...32 Interconnection schemes 6 2 UCClite DC motor and tacho connection...
Страница 35: ...Interconnection schemes 33 6 3 Dual SPA2 DC motor and tacho configuration...