MagCore HF48 Service Manual
Revised: February,2014
Ver. 2014-2
8. Move the Piston module to the position of the calibration tool in Operating area A
again.
9. Turn on the power.
10.
Select the Y tab in the engineering software → (10a) press HOME button.
11. Now the Y-axis will move up slowly to measure the distance.
12. Please record the measured values after the measurement is finished.
13.
Select the X tab in the engineering software → (13a) press HOME button.
14. Now the X-axis will move up slowly to measure the distance.
15. Please record the measured values after the measurement is finished.
16. Cut off the power and repeat above procedures for Operating area B.
10a
10
13a
13