Replace X-axis servo motor
1. Please make sure the instrument is turned off.
2. Disconnect the X-axis servo motor connection and the sensor connection.
3. Detach the X-axis belt. Screw off the 4 screws (5mm hex wrench) on the front
barrier of the driving wheel and the 2 pushing screws (6mm hex wrench)