22
RBCBioscience
Issued By: RBC Bioscience Corp.
MagCore HF48 Service Manual
Revised: February,2014
Ver. 2014-2
Select UV Sterilization in the Maintenance menu.
You may select to run sterilization for 30min or 60min. Press START.
The screen will show the remaining time.
Please press Stop to stop if you want to stop sterilization.
Maintenance
Please select Item
Back
UV Sterilization
Moving Test
Heater Test
Rejection Tip
Piston Test
O-Ring & Piercing Needles
DC Motor Operate
Update Program Files
Start
Stop
30min
60min
UV Sterilization
select Item and press START to run
Back
30:00
Remaining Time
Complete
Back
UV Sterilization
Open the top cover of the Part Chamber.
Replace the old UV lamp with a new one.
Select Status in the main menu and press UV Time Reset key to
reset the hours of UV lamp.
Please test to see if the UV lamp is normal after replacement.
Replace UV lamp
Next Step
Next Step
Next Step
Please note that due to its weight,
the top cover of the Part Chamber
requires two people to lift it.
Status
Back
System Date : 2012.12.12
System Time : 12:12:12
UV Lamp Operating Time : Hr:00 Min:00 Sec:00
UI Software Ver. : Ver.SUPER2012-0.1
PLC Software Ver. : Ver.SUPER2012-0.1
Machine serial number :
Error history list
UV Time Reset