7. Use a thickness gauge to measure the interval between the costal plate of the
Cartridge Rack and the X-axis calibration tools. The interval should be lower than
0.2mm (try 0.1 mm thickness first and then try 0.2mm thickness). The interval
needed by Operating area B is the same as area A; please repeat Step 7 to measure
the interval in Operating area B.
If the tolerance range is exceeded, please adjust it according to the following steps.
1. Support the top cover from its back.
2. Remove the top screws off the two nitrogen cylinders and detach them from the top
cover.
17
Due to the heavy weight of the top cover, multiple personnels may be
required for this step.
Caution !
4. 再將與儀器連接端兩顆螺絲拆下。
Adjust the interval between Y axis and costal plate.