RBCBioscience
Issued By: RBC Bioscience Corp.
MagCore HF48 Service Manual
Revised: February,2014
Ver. 2014-2
3. Please use 30mm wrench to slightly loosen the horizontal bolts and adjust the top
screws (Please fine tune them, you can use a rubber hammer to lightly beat the
monkey wrench) until it conforms to the tolerance range. Fasten the screws.
It will not be necessary to perform this unless the instrument
undergoes serious collision, toppling, inappropriate disassembly,
etc. With the attempt to repair the instrument under such
circumstances through the above procedures, we cannot ensure you
that it can be completely recovered via fine tuning.
Caution !
17
17
Loosen it but don't
completely detach it
from the machine.
Fine tune