Please perform the operation following the steps below.
Test home value
Measure the interval between Y axis and costal plate
After installation, it is necessary to make sure that the position of modules and axes
have not been shifted during shipping. Please check if the values conform to the given
values in the original factory document.
1. Please make sure that the instrument is turned
off.
2. Detach the connection on the brake and install
the electronic brake thread.
3. Install top calibration tool.