39
RBCBioscience
Issued By: RBC Bioscience Corp.
MagCore HF48 Service Manual
Revised: February,2014
Ver. 2014-2
Measure the level height of X-axis and costal plates at
operating area A and operating area B
1. Fasten the X-axis calibration tool at the front side of the Piston module with the
screws from the cover.
2. Move the X-axis calibration tool to the edge of the costal plate on Cartridge Rack of
Operating area A. Measure the distance between the front edge of the costal plate
and the Z-axis calibration tool by an electronic caliper (0.01 accuracy) and record it.
Then move the X-axis calibration tool to the edge of the costal plate on Cartridge
Rack of Operating area B. Measure the distance between the front edge of the costal
plate and the Z-axis calibration tool by the electronic caliper (0.01 accuracy). Make
sure that the error range for the two measured values is within ±0.2mm. Please refer
to the next chapter for lowering tolerance.
Area A
Area B