MagCore HF48 Service Manual
Revised: February,2014
Ver. 2014-2
Replace Y-axis belt
1. Please make sure the instrument is turned off.
2. Disconnect the Y-axis servo motor connection and the sensor connection.
3. Screw off the 4 screws (3mm hex wrench) on the fixing plate of the Y-axis motor.
4. Screw off the 2 screws (2mm hex wrench) on the timing wheel of the Y-axis belt and
move the belt gear up to detach it from the machine.
5. Screw off the 3 screws (4mm hex wrench) above the rack of the Y-axis brake
component.
6. Screw off the 2 screws (2mm hex wrench) of the timing wheel of the Y-axis brake.
7. Attach the belt gear on the Y-axis motor.
8. Install the Y-axis brake component and push it forward to fix the brake and screw
rod. Attach back the rack and the connector.
9. Plug in the power connection of the Y-axis motor and the connections of the sensor.
10. The replacement of the Y-axis motor is finished.