Issued By: RBC Bioscience Corp.
MagCore HF48 Service Manual
Revised: February,2014
Ver. 2014-2
Please place a cartridge into each Cartridge Rack and add 500ul of
water into heat block well 1. Insert a thermometer into heat block
well 1 to measure the temperature.
Select Heater Test in the Maintenance menu.
Press Start. The screen will show the supposed heating temperautre
by calculation. Please compare with the value on your thermometer
to see if they are the same.
Press Stop if you want to stop the test.
Maintenance
Please select Item
Back
UV Sterilization
Moving Test
Heater Test
Rejection Tip
Piston Test
O-Ring & Piercing Needles
DC Motor Operate
Update Program Files
Heater test
Start
Stop
Heater Test
Press START to run
Back
Heater A : 65°C / 148°F
Heater B : 65°C / 148°F
This function is for returning tips to their original locations after
emergency stop or operation interruption due to power outrage.
Select Rejection Tip in the Maintenance menu.
Please select Start to begin. After the process is completed, the
system will return to initial position automatically.
Maintenance
Please select Item
Back
UV Sterilization
Moving Test
Heater Test
Rejection Tip
Piston Test
O-Ring & Piercing Needles
DC Motor Operate
Update Program Files
Rejection Tip
Start
Rejection Tip
Press Start to rejection tip
Back
Rejection B Area Tip
Rejection A Area Tip
A
B
D
A
B
E
Next Step
Next Step
heat block well 1
The acceptable range of temperature is
between 60° C and 70° C.
!