84
RBCBioscience
Issued By: RBC Bioscience Corp.
MagCore HF48 Service Manual
Revised: February,2014
Ver. 2014-2
Test home distance
1.
Please refer to “Installation” → ”Test home value” for the information
about the preparation.
2. Turn on the power.
3. Connect RS232 to USB of the computer and
make sure the connection port number of the
USB wire is COM# in the device manager.
4. Open the engineering software.
5. Change it to the port number on the USB
connection wire and turn on the power. Now
the connection icon should show “connected”.
6.
Select (6a)「AUTO(TAB PAGE)」→「PM operation (item))」→ (6b)「Initinal(Select)」
→ (6c)「PM Start(Button)」→ (6d) confirm「reset P2P command」
7. Turn off the power.
6a
6b
6c
Detach the RS232 wire from
the instrument and connect the
computer to the PLC wire.
Not here
Connected to the
computer
Not connected
Connected
5
6d