Select Moving Test in the Maintenance menu.
Select the axis to test and press Start. The motor will run
continuously until you press Stop.
Select DC Motor Operate in the Maintenance menu.
All consumables are fixed in place when test is stopped. You can
operate the DC motor through this function if you need to raise or
lower the drip pan.
Refer to the page 7 for corresponding locations of operation areas.
Select the area you want to move up or down and press Start. It will
stop automatically when the motor moves to end.
Maintenance
Please select Item
Back
UV Sterilization
Moving Test
Heater Test
Rejection Tip
Piston Test
O-Ring & Piercing Needles
DC Motor Operate
Update Program Files
Maintenance
Please select Item
Back
UV Sterilization
Moving Test
Heater Test
Rejection Tip
Piston Test
O-Ring & Piercing Needles
DC Motor Operate
Update Program Files
Moving Test
Please select Item
Back
X Asix Moveing Test
Y Asix Moveing Test
M Asix Moveing Test
V Asix Moveing Test
ALL Asix Moveing Test
Stop
Start
DC Mortor Operate
Please select Item
Back
Opretion Area A UP
Opretion Area A DOWN
Opretion Area B DOWN
Shutter Close
Opretion Area B UP
Shutter Open
Start
Moving test
Motor elevating operation
Next Step
Next Step
X-axis motor moving test
Motor in operation area A
moves up.
Open Shutter
Motor in operation area B
moves up.
Y-axis motor moving test
Motor in operation area A
moves down.
Close Shutter
Motor in operation area B
moves down.
M-axis motor moving test
V-axis motor moving test
Moving tests for all axes