24
b)
PROD 2 display will now show "Prn" and DATA display will show "End" (Print Field
End).
1)
To Print Field End, depress
.
2)
While "End" is displayed, if Field Begin is required, depress
to toggle DATA display to "BEGn".
c)
Momentarily depress
to advance to FIELD REFERENCE.
3)
FIELD REFERENCE
a)
PROD 1 display will show "SPr". PROD 2 display will show "FrEF". DATA display will show
"0".
b)
Enter the field number.
c)
Momentarily depress
to advance to BAUD RATE.
4)
BAUD RATE
a)
PROD 1 display will show "SPr". PROD 2 display will show "bAUd". DATA display will show
"1200".
b)
Depressing
momentarily changes the DATA display between "1200" and "9600".
c)
Momentarily depress
to advance to RATE CHANGE ALARM ON/OFF.
5)
RATE CHANGE ALARM ON/OFF
a)
PROD 1 display will show "SPr". PROD 2 display will show "rAtE". DATA
display will show "on".
b)
Depressing
momentarily changes the DATA display between "on" and
"oFF". A value of "on" means alarm is enabled; a value of "oFF" means alarm
is disabled.
c)
Momentarily depress
to advance to DATA LOGGER TRIGGER VALUE.
6)
DATA LOGGER TRIGGER VALUE
a)
PROD 1 display will show "SPr". PROD 2 display will show "TriG". DATA display will show
"0".
b)
Enter the TRIGGER VALUE.
c)
Momentarily depress
to advance to DATA LOGGER TRIGGER UNITS.
Содержание SCS 700
Страница 1: ...Installation Service Manual SCS 700 ...
Страница 2: ...W A R N I N G Disconnect console before jump starting charging battery or welding on equipment ...
Страница 3: ......
Страница 4: ... ...
Страница 58: ...54 APPENDIX 9 INJECTION MODULE MOUNTING INSTRUCTIONS 1 TYPICALMODULEDIAGRAM ...
Страница 64: ...60 APPENDIX 14 SCS 710 6 BOOM CONSOLE WIRING DIAGRAM ...
Страница 69: ......
Страница 70: ......