Remote Control
R&S
®
FSPN
329
User Manual 1179.4363.02 ─ 01
Usage:
Query only
Manual operation:
See
FETCh<n>:PNOise<t>:SPURs:DISCrete?
This command queries the discrete jitter result.
Suffix:
<n>
.
<t>
Return values:
<Jitter>
<numeric value>
Default unit: s
Example:
//Query discrete jitter
FETC:PNO:SPUR:DISC?
would return, e.g.
2.3e-08
Usage:
Query only
Manual operation:
See
FETCh<n>:PNOise<t>:SPURs:JITTer?
This command queries the jitter of the measured spurs.
Suffix:
<n>
.
<t>
Return values:
<Jitter>
<numeric value>
Default unit: s
Example:
//Query jitter of spurs
FETC:PNO:SPUR:JITT?
would return, e.g.
+2.199004,+2.440125142E-012,
+7.332004,[...]
Usage:
Query only
Manual operation:
See
FETCh<n>:PNOise<t>:SPURs:RANDom?
This command queries the random jitter result.
Suffix:
<n>
.
Measurements and Result Displays