Data Management
R&S
®
FSPN
195
User Manual 1179.4363.02 ─ 01
If you make any changes to the settings
after
storing the configuration file, remember
to save the settings again. Otherwise those settings cannot be restored and will be
overwritten by the stored values when the configuration file is recalled.
To recall configurable instrument settings
1. Select the "Open" icon from the toolbar.
2. In the "Recall" dialog box, switch to the "Recall" tab.
3. In the file selection dialog box, select the filename and storage location of the set-
tings file.
Note:
The "File Type" indicates whether the file contains instrument settings for
all
"Channel" s, or only those for the current "Channel" .
4. If several items were saved, select which items are restored.
5. If a "Channel" was saved, select whether the settings will replace the settings in the
current "Channel" , or whether a new channel with the saved settings will be
opened.
6. Select "Recall" .
The settings and selected items from the saved measurement are restored and
you can repeat the measurement with the same settings.
Note that any changes made to the settings
after
storing the configuration file will
be overwritten by the stored values when the configuration file is recalled.
To recall settings automatically after preset or reboot
You can define the settings that are restored when you preset or reboot the instrument.
1. Configure the settings as required and save them as described in
2. In the "Save/Recall" menu, select "Startup Recall" .
3. From the file selection dialog box, select the recall settings to restore.
4. Select "Select File" .
5. Set "Startup Recall" to "On" .
Now when you press the [PRESET] key or reboot the instrument, the defined set-
tings will be restored.
6. To restore the factory preset settings, set "Startup Recall" to "Off" .
8.3
Import/Export Functions
Access
: "Save" / "Open" icon in the toolbar > "Import" / "Export"