User Manual 1179.4363.02 ─ 01
5. Enter the command
C:\R_S\INSTR\USER\user\AUTOLOGIN.REG
.
6. Press the [ENTER] key to confirm.
The auto-login function is reactivated with the changed password. It will be applied
the next time the instrument is switched on.
Switching users when using the auto-login function
Which user account is used is defined during login. If auto-login is active, the login win-
dow is not displayed. However, you can switch the user account to be used even when
the auto-login function is active.
1. Select the "Windows" icon in the toolbar to access the operating system of the
R&S
FSPN (see also
2. Press [CTRL] + [ALT] + [DEL], then select "Sign out".
The "Login" dialog box is displayed, in which you can enter the different user
account name and password.
For information on deactivating and reactivating the auto-login function, see
3.1.12
Checking the Supplied Options
The instrument can be equipped with both hardware and firmware options. To check
whether the installed options correspond to the options indicated on the delivery note,
proceed as follows.
1. Press the [SETUP] key.
2. Press the "System Config" softkey.
3. Switch to the "Ve Options" tab in the "System Configuration" dialog box.
A list with hardware and firmware information is displayed.
4. Check the availability of the hardware options as indicated in the delivery note.
3.1.13
Performing a Self-Alignment
When strong temperature changes occur in the environment of the R&S
FSPN, or after
updating the firmware, you have to perform a self-alignment to align the data to a refer-
ence source.
During self-alignment, do not connect a signal to the RF input connector. Running a
self-alignment with a signal connected to the RF input can lead to false measurement
results.
Performing a self-alignment
Before performing this functional test, make sure that the instrument has reached its
operating temperature (for details, refer to the data sheet).
Preparing for Use