Common Analysis and Display Functions
R&S
®
FSPN
182
User Manual 1179.4363.02 ─ 01
All stored limit lines with the file extension
.LIN
in the
limits
subfolder of the
main installation folder of the instrument that are compatible to the current mea-
surement settings are displayed in the overview.
How to activate and deactivate a limit check
A limit check is automatically performed as soon as any of the limit lines is activated.
1. To activate a limit check:
Select the "Check Traces" setting for a limit line in the overview and select the
trace numbers to be included in the limit check. One limit line can be assigned to
several traces.
The specified traces to be checked are compared with the active limit lines. The
status of the limit check is indicated in the diagram.
2. To deactivate a limit line, deactivate all "Traces to be Checked" for it.
To deactivate all limit lines at once, select the "Disable All Lines" button.
The limit checks for the deactivated limit lines are stopped and the results are
removed form the display.
How to edit existing limit lines
Existing limit line configurations can be edited.
1. In the "Line Config" dialog box, select the limit line.
2. Select the "Edit" button.
3. Edit the line configuration as described in
"How to configure a new limit line"
4. Save the new configuration by selecting the "Save" button.
If the limit line is active, the edited limit line is displayed in the diagram.
How to copy an existing limit line
1. In the dialog box, select the limit line.
2. Select the "Line Config" "Copy To" button.
3. Define a new name to create a new limit with the same configuration as the source
line.
4. Edit the line configuration as described in
"How to configure a new limit line"
5. Save the new configuration by selecting the "Save" button.
The new limit line is displayed in the overview and can be activated.
How to delete an existing limit line
1. In the "Line Config" dialog box, select the limit line.
Limit Lines