Remote Control
R&S
®
FSPN
485
User Manual 1179.4363.02 ─ 01
DIAGnostic:SERVice:INPut[:SELect]
<Signal>
This command activates or deactivates the use of an internal calibration signal as input
for the R&S
FSPN.
Parameters:
<Signal>
CALibration
Uses the calibration signal as RF input.
MCALibration
Uses the calibration signal for the microwave range as RF input.
RF
Uses the signal from the RF input.
*RST:
RF
Example:
DIAG:SERV:INP CAL
Uses the calibration signal as RF input.
Manual operation:
See
See
See
See
DIAGnostic:SERVice:INPut:SYNThtwo[:FREQuency]
<Frequency>
This command selects the frequency which the synthesizers are calibrated for.
The command is available when you select the synthesizer as the calibration source
with
DIAGnostic:SERVice:INPut[:SELect]
Parameters:
<Frequency>
Default unit: Hz
Example:
DIAG:SERV:INP:SEL SYNT
DIAG:SERV:INP:SYNT 10MHZ
Manual operation:
See
DIAGnostic:SERVice:STESt:RESult?
This command queries the self-test results.
Return values:
<Results>
String of data containing the results.
The rows of the self-test result table are separated by commas.
Example:
DIAG:SERV:STES:RES?
would return, e.g.
"Total Selftest Status:
PASSED", "Date (dd/mm/yyyy): 09/07/2004 TIME:
16:24:54", "Runtime: 00:06", "...
Usage:
Query only
Configuring the R&S
FSPN