Operating Manual 1302.6163.12 - 03
4.41
R&S ESU
Receiver Mode
Structure of the ASCII file
Table 4-1
RECEIVER mode, final measurement data – File header
Table 4-2
RECEIVER mode, final measurement data – Data section of the file
Content of file
Description
Type; R&S ESU;
Instrument model
Version;2.32;
Firmware version
Date;03.Aug 2004;
Date record storage date
Mode;Receiver;
Instrument operating mode
Start;10000;Hz
Stop;100000;Hz
Start/stop of the display range.
Unit: Hz
x-Axis;LIN;
Scaling of x axis linear (LIN) or logarithmic (LOG)
Scan Count;1;
Number of scans set
Transducer;TRD1;
Transducer name (if switched on)
Scan 1:
Loop over all defined scan ranges (1-10)
Start;150000;Hz
Range - start frequency in Hz
Stop;1000000;Hz
Range – stop frequency in Hz
Step;4000;Hz
Range - step width in Hz for linear step width or in% (1-100)
for logarithmic step width
RBW;100000;Hz
Range - resolution bandwidth
Meas Time;0.01;s
Range - measurement time
Auto Ranging;ON;
Auto ranging on - or off for current range
RF Att;20;dB
Range - input attenuation
Auto Preamp;OFF;
Auto Preamp on or off for current range
Preamp;0;dB
Range - preamplifier on (20dB) or off (0dB)
Content of file
Description
TRACE 1 FINAL:
Selected trace
Trace Mode;CLR/WRITE;
Trace mode: CLR/WRITE,AVERAGE,MAX HOLD,MIN
HOLD, VIEW, BLANK
Final Detector, QUASIPEAK;
Final detector MAX PEAK, MIN PEAK, RMS, AVERAGE,
QUASIPEAK, CISPR RMS, CISPR AV
x-Unit;Hz;
Unit of x values:
y-Unit;dBuV;
Unit of y values:
Final Meas Time;1.000000;s
Final measurement time
Margin;6.000000:s
Margin
Values;8;
Number of test points
2;154000.000000;81.638535;15.638
535;N;GND
1;158000.000000;86.563789;7.5637
89;N;GND
2;1018000.000000;58.689873;-
1.310127;N;GND
...
Measured values:
<Trace>;<x-value>, <y-value>; <phase>; <ground>
Phase and protective grounding are output only if a line
impedance stabilization has been activated. They specify the
setting at which the maximum RFI level at the associated
frequency was found.
Содержание 1129.7246.03
Страница 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual ...
Страница 25: ......
Страница 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03 ...
Страница 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03 ...
Страница 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03 ...
Страница 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03 ...
Страница 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03 ...
Страница 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03 ...
Страница 821: ...R S ESU Remote Control Programming Examples More Complex Programming Examples 7 32 Operating Manual 1302 6163 12 03 ...
Страница 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03 ...
Страница 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU ...
Страница 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03 ...