R&S ESU
Remote Control – Description of Commands
CALCulate Subsystem
6.34
Operating Manual 1302.6163.12 - 03
6.5.3.2
CALCulate:LIMit:CONTrol Subsystem
The CALCulate:LIMit:CONTrol subsystem defines the x-axis (CONTrol-axis).
CALCulate<1|2>:LIMit<1...8>:CONTrol[:DATA]
<numeric_value>,<numeric_value>..
This command defines the x-axis values (frequencies or times) of the upper or lower limit lines.
The values are defined independently of the measurement window.
The number of values for the CONTrol axis and for the corresponding UPPer and/or LOWer limit lines
have to be identical. Otherwise default values are entered for missing values or not required values
are deleted.
In analyzer mode, the unit of values depends on the frequency or time domain of the x-axis, i.e. it is
HZ
with
CALC:LIM:CONT:DOM FREQ and S with CALC:LIM:CONT:DOM TIME.
Example:
"CALC:LIM2:CONT 1MHz,30MHz,100MHz, 300MHz,1GHz"
'Defines 5 reference values for the x-axis of limit line 2
"CALC:LIM2:CONT?"
'Outputs the reference values for the x-axis of limit line 2 separated by a comma.
Characteristics:
*RST value: - (LIMit:STATe is set to OFF)
SCPI: conform
Mode:
R, A
CALCulate<1|2>:LIMit<1...8>:CONTrol:DOMain
FREQuency | TIME
This command defines the frequency or time domain for the x-axis values.
Example:
"CALC:LIM2:CONT:DOM TIME"
'Defines the time domain for the x-axis of limit line 2.
Characteristics:
*RST value: FREQuency
SCPI: device-specific
Mode:
R, A
CALCulate<1|2>:LIMit<1...8>:CONTrol:OFFSet
<numeric_value>
This command defines an offset for the x-axis value of the selected relative limit line in the frequency
or time domain.
The unit of values depends on the frequency or time domain of the x-axis, i.e. it is
HZ
with
CALC:LIM:
CONT:DOM FREQ and S with CALC:LIM:CONT:DOM TIME.
Example:
"CALC:LIM2:CONT:OFFS 100us"
'Sets the X offset for limit line 2 (defined in the time domain) to 100µs.
Characteristics:
*RST value: 0
SCPI: device-specific
Mode:
R, A
Содержание 1129.7246.03
Страница 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual ...
Страница 25: ......
Страница 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03 ...
Страница 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03 ...
Страница 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03 ...
Страница 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03 ...
Страница 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03 ...
Страница 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03 ...
Страница 821: ...R S ESU Remote Control Programming Examples More Complex Programming Examples 7 32 Operating Manual 1302 6163 12 03 ...
Страница 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03 ...
Страница 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU ...
Страница 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03 ...