Operating Manual 1302.6163.12 - 03
4.93
R&S ESU
IF Spectrum Analysis Mode
VIEW
The
VIEW
softkey freezes the current contents of the trace memory and displays it.
For further information refer to the descriptions in section
–
.
BLANK
The
BLANK
softkey activates the blanking of the trace on the screen.
Remote command:
DISP:WIND:TRAC OFF
MIN HOLD
The
MIN HOLD
softkey activates the min hold mode for the trace. R&S ESU saves
for each scan the smallest of the previously stored/currently measured values in the
trace memory.
This function is useful e.g. for making an unmodulated carrier in a composite signal
visible. Noise, interference signals or modulated signals are suppressed by the min
hold function whereas a CW signal is recognized by its constant level.
Remote command:
DISP:WIND:TRAC:MODE MINH
AVG MODE
LOG | LIN
The
AVG MODE LOG/LIN
softkey selects logarithmic or linear averaging for the log-
arithmic level display mode.
Remote command:
CALC:MATH:AVER:MODE LIN
With logarithmic averaging, the dB values of the display voltage are averaged. With
linear averaging the level values in dB are converted into linear voltages or powers
prior to averaging. Voltage or power values are averaged or offset against each
other and reconverted into level values.
For stationary signals the two methods yield the same result.
Logarithmic averaging is recommended if sinewave signals are to be clearly visible
against noise since with this type of averaging noise suppression is improved while
the sinewave signals remain unchanged.
For noise or pseudo-noise signals the positive peak amplitudes are decreased in
logarithmic averaging due the characteristic involved and the negative peak values
are increased relative to the average value. If the distorted amplitude distribution is
averaged, a value is obtained that is smaller than the actual average value. The dif-
ference is -2.5 dB.
2 .5 d B
A m p litu d e
P r o b a b ilit y d is tr ib u tio n
A m p litu d e d istr ib u tio n
( a fte r a v e ra g in g )
A m p litu d e d istr ib u tio n
( w ith o u t a v e ra g in g )
Содержание 1129.7246.03
Страница 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual ...
Страница 25: ......
Страница 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03 ...
Страница 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03 ...
Страница 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03 ...
Страница 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03 ...
Страница 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03 ...
Страница 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03 ...
Страница 821: ...R S ESU Remote Control Programming Examples More Complex Programming Examples 7 32 Operating Manual 1302 6163 12 03 ...
Страница 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03 ...
Страница 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU ...
Страница 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03 ...