R&S ESU
Remote Control – Description of Commands
SENSe Subsystem
6.164
Operating Manual 1302.6163.12 - 03
[SENSe<1|2>:]CORRection:TSET:BREak
ON | OFF
This command defines if the sweep is to be stopped on changeover from range to another. Prior to this
command, the command
SENS:CORR:TSET:SEL
must be sent.
Example:
"CORR:TSET:SEL 'SET1'"
"CORR:TSET:BRE ON"
Characteristics:
*RST value: OFF
SCPI: device-specific
Mode:
R, A
[SENSe<1|2>:]CORRection:TSET:CATalog?
This command reads out the names of all transducer sets stored on the internal drive.
Syntax of output format:
<Sum of file lengths of all subsequent files>,<free memory on hard disk>, <1st file
name>,<1st file length>, <2nd file name>,<2nd file length>,....,<nth file name>, <nth file length>
Example:
"CORR:TSET:SEL 'SET1'"
"
CORR:TSET:CAT?"
Characteristics:
*RST value: --
SCPI: device-specific
Mode:
R, A
[SENSe<1|2>:]CORRection:TSET:COMMent
<string>
This command defines the comment for the selected transducer set. Prior to this command, the
command
SENS:CORR:TSET:SEL
must be sent.
Example:
"CORR:TSET:SEL 'SET1'"
"
CORR:TSET:COMM 'SET FOR ANTENNA'"
Characteristics:
*RST value: '' (empty comment)
SCPI: device-specific
Mode:
R,A
[SENSe<1|2>:]CORRection:TSET:RANGe<1 to 10>
<freq>,<freq>,<name>..
This command defines a partial range of the selected transducer set. The partial range is determined
by its start and stop frequencies plus a list of names of the assigned transducer factors. The ranges 1
to 10 must be sent in ascending order. Prior to this command, the command
SENS:CORR:TSET:SEL
must be sent.
Parameter:
<freq>,<freq>::= start frequency, stop frequency of the range
<name>...::= list of names for the assigned transducer factors.
The individual names must be characterized by single quotation marks (') and
separated by commas.
Example:
"CORR:TSET:SEL 'SET1'"
"CORR:TRAN:TSET:RANG 1MHZ,2MHZ,'FACTOR1,'FACTOR2'"
Characteristics:
*RST value: -
SCPI: device-specific
Mode:
R, A
Содержание 1129.7246.03
Страница 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual ...
Страница 25: ......
Страница 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03 ...
Страница 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03 ...
Страница 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03 ...
Страница 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03 ...
Страница 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03 ...
Страница 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03 ...
Страница 821: ...R S ESU Remote Control Programming Examples More Complex Programming Examples 7 32 Operating Manual 1302 6163 12 03 ...
Страница 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03 ...
Страница 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU ...
Страница 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03 ...