R&S ESU
Instrument Functions
Analyzer Mode
4.222
Operating Manual 1302.6163.12 - 03
Table 4-1
Example: ASCII export table - file header
The data section for the measurement values starts with the key word "TRACE <n>:
", where <n> includes the number of the used trace. Next comes the peak list with
several columns also separated by ';'.
Table 4-2
Example: ASCII export table - file data section
Spreadsheet programs such as MS Excel can read this format. Use ';' as the delim-
iter for the table cells.
Remote command:
MMEM:STOR:SPUR,'F:\TEST.ASC'
File contents
Explanation
Type;R&S ESU;
Model
Version;3.90;
Firmware version
Date;02.Aug 2006;
Storage date of data set
Mode;ANALYZER; SPURIOUS;
Operating mode of the device
Start;9000.000000;Hz
Stop;7000000000.000000;Hz
Start/stop of the span
Unit: Hz
x-Axis;LIN;
Linear (LIN) or logarithmic (LOG) (future feature)
scaling of the x-axis
Sweep Count;1;
Selected number of sweeps
File contents
Explanation
TRACE 1:
Selected trace
Trace Mode;CLR/WRITE;
Display mode of trace:
CLR/WRITE,AVERAGE,MAX HOLD,MIN HOLD,
VIEW, BLANK
x-Unit;Hz;
Unit of x values:
y-Unit;dBm;
Unit of y values:
Margin;6.000000:s
Peak List margin
Values;8;
Number of measurement points
1;1548807257.5999999000;-
65.602280;-5.602280
1;1587207214.4000001000;-
65.327530;-5.327530
1;2112006624.0000000000;-
4.388008;55.611992
Measurement values:
<Trace>;<x value>; <y value>;<delta limit>
Different language versions of evaluation programs may require different handling
of the decimal point. Using the
DECIM SEP
softkey, you can thus choose between
the delimiters '.' (decimal point) and ',' (comma).
Содержание 1129.7246.03
Страница 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual ...
Страница 25: ......
Страница 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03 ...
Страница 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03 ...
Страница 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03 ...
Страница 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03 ...
Страница 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03 ...
Страница 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03 ...
Страница 821: ...R S ESU Remote Control Programming Examples More Complex Programming Examples 7 32 Operating Manual 1302 6163 12 03 ...
Страница 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03 ...
Страница 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU ...
Страница 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03 ...