R&S ESU
Receiver Mode
4.30
Operating Manual 1302.6163.12 - 03
MIN PEAK
The
MIN PEAK
softkey activates the min peak detector.
Remote command:
DET:REC NEG
QUASIPEAK
The
QUASIPEAK
softkey selects the quasipeak detector.
The IF bandwidth is adapted as a function of the frequency range. The coupling of
the IF bandwidth to the frequency range can be cancelled using the softkey
CISPR
RBW UNCOUPLED
.
Remote command:
DET:REC QPE
AVERAGE
The
AVERAGE
softkey activates the average detector.
Remote command:
DET:REC AVER
CISPR
AVERAGE
The
CISPR AVERAGE
softkey activates the weighting average detector according
to CISPR 16-1. The IF bandwidth is automatically set to the required value accord-
ing to the receiving frequency. This coupling can be cancelled by the softkey
CISPR
RBW UNCOUPLED
.
Remote command:
DET:REC CAV
RMS
The
RMS
softkey activates the rms detector.
Remote command:
DET:REC RMS
CISPR
RMS
The
CISPR RMS
softkey activates the weighting rms detector according to CISPR
16-1. The IF bandwidth is automatically set to the required value according to the
receiving frequency. This coupling can be cancelled by the
CISPR RBW UNCOU-
PLED
softkey.
Remote command:
DET:REC CRMS
CISPR RBW
UNCOUPLED
The
CISPR RBW UNCOUPLED
softkey cancels the coupling of the IF bandwidth to
the frequency range with activated quasipeak or CISPR average detector.
Remote command:
BAND:AUTO OFF
4.5.4.2
Setting the Measurement Time
The measurement time can be set with a 2-digit resolution in the range 10 µs to
100 s.
The measurement time is the time during which R&S ESU measures the input sig-
nal and forms a measurement result weighted by the selected detector. The mea-
surement time does not include settling times of the synthesizer and the IF filter.
R&S ESU automatically waits until transients are over.
When the quasipeak detector is used, the minimum measurement time is 0.5 ms.
When the CISPR average detector or the CISPR rms detector is used, the minimum
measurement time is 100 ms. With the average, RMS, or min/max peak detector the
smallest settable measurement time depends on the bandwidth.
Содержание 1129.7246.03
Страница 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual ...
Страница 25: ......
Страница 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03 ...
Страница 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03 ...
Страница 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03 ...
Страница 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03 ...
Страница 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03 ...
Страница 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03 ...
Страница 821: ...R S ESU Remote Control Programming Examples More Complex Programming Examples 7 32 Operating Manual 1302 6163 12 03 ...
Страница 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03 ...
Страница 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU ...
Страница 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03 ...