PROGRAMMING
4-18
4.5.1
Change Low Alarm Limit
The second sub-menu item in the Alarm Limit sub-menu
allows the user to change the Low Alarm limit. The LCD
displays “Low limit?” To change Low Alarm limit, press
[Y/+]
key, or Press
[N/-]
key advance to next sub-menu in
Table 4.5.
1.
Press
[Y/+]
and the display will show a flashing cursor
on the left-most digit of the previously stored Low
alarm limit.
2.
To modify this limit value, use the
[Y/+]
or
[N/-]
key
to change the digit value and press the
[MODE]
key
momentarily to advance to the next digit. The flashing
digit will move to the next digit to its right. Repeat
this process until the new limit values is entered. Press
and hold the
[MODE]
key for 1 second to exit data
entry mode.
3.
If there is any change to the existing value, the display
shows “Save?” Press
[Y/+]
to accept the new value
and move to the next sub-menu. Press
[N/-]
to discard
the changes and move to the next sub-menu.