![Rae MiniRAE 2000 Скачать руководство пользователя страница 137](http://html1.mh-extra.com/html/rae/minirae-2000/minirae-2000_operation-and-maintenance-manual_3377883137.webp)
TROUBLESHOOTING
8-10
8.2 Troubleshooting Table
Table 8-2
Problem
Possible Reasons & Solutions
Cannot turn on
power after
charging the
battery
Reasons:
Discharged battery.
Defective battery.
Microcomputer hang-up.
Solutions:
Charge or replace battery.
Disconnect, then connect battery
to reset computer.
No LCD back
light
Reasons:
Trigger level too low, the
current mode is not user mode,
and the mode does not support
automatic turn on back light.
Solutions:
Adjust trigger level.
Verify the back light can be
turned on in user mode.
Call authorized service center.
Lost password
Solutions:
Use default configuration to
reset password from computer.
Warning : If the default
configuration is loaded, all the
parameters need to be reset,
including pump, lamp and
LCD contrast thresholds,
serial number; and re-
calibration is required.
Reading
abnormally
High
Reasons:
Dirty sensor module.
Dirty water trap filter.
Excessive moisture and water
condensation.
Solutions:
Clean sensor module and lamp
housing.
Replace water trap filter.
Blow dry the sensor module.