COMPUTER INTERFACE
5-34
5.8
Upgrade the Firmware
To upgrade the firmware in the MiniRAE 2000 monitor,
open the ProRAE-Suite software. Do not open any
configuration file or data file so the ProRAE-Suite
software has no client window open, as shown in Figure
5.14. Click the
Option -> Load Firmware
menu item; a
message box will appear to remind the user to connect the
instrument to the PC through the serial port. After making
the proper connection, click the
OK
button. An
Open
dialog box is displayed, as shown in Figure 5-16. Select
the firmware file (.a07) to be loaded into the instrument,
and click the
Open
button to start the loading process.
After the firmware loading is finished, a message box will
appear on the screen to let the user know that the
firmware is successfully loaded into the instrument.
Figure 5-16 Open Firmware File Dialog Box