GENERAL INFORMATION
1-2
1.1 General Specifications
Table 1.1
Portable VOC Monitor Specification
Size:
8.2"L x 3.0"W x 2.0"H
Weight:
19.5 oz with battery pack
Detector:
Photo-ionization sensor with 9.8, 10.6,
or 11.7 eV UV lamp
Battery:
A 4.8V /1250 mAH Rechargeable
Nickel Metal Hydride battery pack
(snap in, field replaceable)
Battery Charging: 10 hours charge through built-in
charger
Operating Hours:
Up to10 hours continuous operation
Display:
1 line by 8 characters 5x7 dot matrix
LCD (0.4” character height) with LED
back light automatically in dim light
Range, Resolution & Response time (t90):
Isobutylene (calibration gas)
0-99 ppm
0.1 ppm
2 sec
100-1,999 ppm 1.0 ppm
2 sec
2000-10,000 ppm 1.0 ppm 2 sec
Measurement Accuracy (Isobutylene):
0 – 2000 ppm:
±
2 ppm or 10% of
reading.
> 2000 ppm:
±
20% of reading
PID Detector:
Easy access to lamp and sensor for
cleaning and replacement
Correction Factors: Built-in 102 VOC gases