Maintenance and Performance Checks 5-11
4.
Set Pulse/Function generator to output 2 Vp-p square
waveforms. Change 3152 and Pulse/Function generator
frequency and verify counter phase readings as shown
below.
3152 FREQ.
PULSE GENERATOR
SETTING
FREQ. SETTING
COUNTER READING
10.00 MHz
10.00 MHz
0
E
±77
E
1.000 MHz
1.000 MHz
0
E
±12.2
E
100.0 KHz
100.0 KHz
0
E
±5.72
E
10.00 KHz
10.00 KHz
0
E
±5
E
5.
Change the Model 3152 PHAS2:ADJ setting to 180
E
.
Change 3152 and Pulse/Function generator frequency and
verify counter phase readings as shown below.
3152 FREQ.
PULSE GENERATOR
SETTING
FREQ. SETTING
COUNTER READING
10.00 KHz
10.00 KHz
180
E
±5
E
6.
Change the Model 3152 PHAS2:ADJ setting to -180
E
.
Change 3152 and Pulse/Function generator frequency and
verify counter phase readings as shown below.
3152 FREQ.
PULSE GENERATOR
SETTING
FREQ. SETTING
COUNTER READING
10.00 KHz
10.00 KHz
-180
E
±5
E
7.
Change the Model 3152 PHAS2:ADJ setting to 0
E
.
Change 3152 and Pulse/Function generator frequency and
verify counter phase readings as shown below. Note the
exact phase reading.
3152 FREQ.
PULSE GENERATOR
SETTING
FREQ. SETTING
COUNTER READING
10.00 KHz
10.00 KHz
0
E
±5
E
(Note and
record
the exact reading)
Содержание 3152
Страница 16: ...Getting Started 1 7 Figure 1 1 Segment 1 Sin x x Waveform Figure 1 2 Segment 2 Sine Waveform...
Страница 25: ...Configuring The Instrument 2 6 This page was intentionally left blank...
Страница 63: ...Using The Instrument 3 38 This page was intentionally left blank...
Страница 80: ...SCPI Command Reference 4 17 Figure 4 1 SCPI Status Registers...
Страница 121: ...Specifications A 12 This page was intentionally left blank...