Maintenance and Performance Checks 5-5
3. Connect the Model 3152 output to the counter’s input.
Change the frequency setting as required for the test and
verify the reading on the counter display as follows:
3152 SETTING
COUNTER READING
10.00000 Hz 9.999000 Hz - 10.00100 Hz
100.0000 Hz 99.99000 Hz - 100.0100 Hz
1.000000 KHz
999.9000 Hz - 1.000100 KHz
10.00000 KHz
9.999000 KHz - 10.00100 KHz
100.0000 KHz
99.99000 KHz - 100.0100 KHz
1.000000 KHz
999.9000 KHz - 1.000100 KHz
10.00000 MHz
9.999000 MHz - 10.00100 MHz
50.00000 MHz
49.99500 MHz - 50.00500 MHz
Amplitude Accuracy
Accuracy Specifications (1 KHz):
±(1% of r200 µV), to 160 mV
±(1% of r2 mV), to 1.6 V
±(1% of r20 mV), to 16 V
Equipment: DMM, 50
S
feedthrough termination
1.
Configure the Model 3152 as follows:
Function:
Sine
Frequency:
1.000 KHz
Amplitude:
15 V
Remote Commands: :RES
:OUTP ON
:FUNC:SHAP SIN
:FREQ 1000
:VOLT 15
2.
Set the DMM to ACV measurements (RMS).
3.
Connect the Model 3152 output to DMM input. Terminate
the output with a 50
S
feedthrough termination. Set the
amplitude and verify DMM reading as follows:
3152 SETTING
DMM READING
15.00 V
5.373 V - 5.230 V
1.500 V
537.3 mV - 523.0 mV
150.0 mV
53.73 mV - 52.30 mV
DC Offset
Characteristics
Accuracy Specifications:
±(1% of r1% of amp200 µV), ±80 mV window
±(1% of r1% of amp2 mV), ±800 mV window
±(1% of r1% of amp20 mV) ±8 V window
Содержание 3152
Страница 16: ...Getting Started 1 7 Figure 1 1 Segment 1 Sin x x Waveform Figure 1 2 Segment 2 Sine Waveform...
Страница 25: ...Configuring The Instrument 2 6 This page was intentionally left blank...
Страница 63: ...Using The Instrument 3 38 This page was intentionally left blank...
Страница 80: ...SCPI Command Reference 4 17 Figure 4 1 SCPI Status Registers...
Страница 121: ...Specifications A 12 This page was intentionally left blank...